VKC ScienceChemistry for written exams Telegram

UPSC Combined Geo-Scientist

Chemist

  • 3 descriptive papers
  • 200 marks each
  • 3 h per paper

Mains Paper III covers analytical and organic chemistry.

UPSC Combined Geo-Scientist, Mains · 2017 · Mains Paper III

Q1(f). Flame ionisation detector

Question

Explain the FID and name two compounds that it cannot normally measure.

Answer in brief

FID measures the ion current generated when organic compounds burn in a hydrogen–air flame. CO₂ and H₂O give essentially no useful response.

Key points

  • Column effluent enters the flame; carbon-containing fragments generate charged species during combustion.
  • A potential between burner and collector draws the ions to an electrode. The resulting small current is amplified and plotted against retention time.
  • Response is approximately proportional to the mass flow of reduced organic carbon over a broad range. It is destructive and very sensitive for hydrocarbons.
  • Fully oxidised carbon compounds such as CO₂ and inorganic substances such as water have negligible normal FID response.

The full model answer

As written for the answer booklet: the diagram carries the working, the lines stay short. Read here only; it is not offered as a download.

Model answer to Q1(f), 2017 Mains Paper III: solution pages 3 and 4
Solution pages 3 and 4 of 18

CoversFlame ionisation detectorGas chromatography