Q1(e). Properties needed in a detector for X-ray fluorescence

Question

What are the properties vital for detectors to be used in X-ray fluorescence instrument?

Answer in brief

High detection efficiency over the X-ray range used, pulse height proportional to photon energy, good energy resolution, short dead time (linear response at high count rates), low noise and long-term stability.

Key points

  • High detection (quantum) efficiency over the range used (about 1–40 keV); a thin, low-Z window (Be, polymer) admits soft X-rays of light elements.
  • Proportionality: pulse height ∝ photon energy, allowing pulse-height selection (WDXRF) and energy dispersion (EDXRF).
  • Good energy resolution (small FWHM): Si(Li) or silicon drift detector about 130–150 eV at Mn Kα (5.9 keV).
  • Short dead time and linear response at high count rates: dead time about 1 µs or less.
  • Low noise and low background; semiconductor detectors are cooled (liquid N₂ or Peltier) to lower thermal noise.