Q1(f). X-ray fluorescence in X-ray diffraction

  • 5 marks
  • Topic: X-ray methods
  • Answered
Question

What is the cause of fluorescence in X-ray diffraction? How does fluorescence phenomenon impact X-ray diffraction data?

Answer in brief

When the incident photon energy exceeds the K absorption edge of an element in the sample (Fe, Co, Mn with Cu Kα), core electrons are ejected and outer electrons fall back, emitting characteristic fluorescent X-rays in all directions. They raise the background and lower the peak-to-background ratio.

Key points

  • Cu Kα (8.04 keV) lies above the K edges of Fe (7.11 keV), Co (7.71 keV) and Mn (6.54 keV). Photoionisation, not diffraction, is absorbed; the Kα photon emitted by Fe has a random direction.
  • (i) high, sloping background; (ii) weaker peak/background, noisy weak peaks, poorer detection limit; (iii) strong absorption (high mass absorption coefficient) so the intensity from the bulk falls. Peak positions are not changed.
  • use Co Kα or Cr Kα for Fe samples; a diffracted-beam monochromator or an energy-discriminating (Si drift) detector; a thin Ni or Fe filter.

The full model answer

Worked out step by step, as it should go in the answer booklet. The solution sheet follows it.

  • Cause: Cu Kα (8.04 keV) lies above the K edges of Fe (7.11 keV), Co (7.71 keV) and Mn (6.54 keV). Photoionisation, not diffraction, is absorbed; the Kα photon emitted by Fe has a random direction.
  • Effect: (i) high, sloping background; (ii) weaker peak/background, noisy weak peaks, poorer detection limit; (iii) strong absorption (high mass absorption coefficient) so the intensity from the bulk falls. Peak positions are not changed.
  • Remedy: use Co Kα or Cr Kα for Fe samples; a diffracted-beam monochromator or an energy-discriminating (Si drift) detector; a thin Ni or Fe filter.
Model answer to Q1(f), 2026 Mains Paper III: solution page 6
Solution page 6 of 30

CoversXRD backgroundX-ray fluorescence in X-ray diffraction