UPSC Combined Geo-Scientist, Mains · 2026 · Mains Paper III
Q1(g). Argon in ICP-MS; acids for arsenic
Question
(i) What is the purpose of using argon gas in ICP-MS? (ii) Why is HCl or HClO₄ not recommended for sample preparation for arsenic analysis by ICP-MS?
Answer in brief
(i) Argon (IE 15.76 eV) forms the plasma (6000–10 000 K) that atomises and ionises the sample and acts as nebuliser and coolant gas; it is inert. (ii) ⁴⁰Ar³⁵Cl⁺ has m/z=75, the same as ⁷⁵As⁺: a polyatomic interference; HCl and HClO₄ supply the chlorine.
Key points
- (i) 2 marks: the plasma gas of the torch; its ionisation energy (15.76 eV) exceeds the first IE of nearly all elements (not F, Cl, Ne, He), so analytes give singly charged ions; it is inert, giving no oxide or compound lines;…
- (ii) 3 marks: arsenic is monoisotopic, ⁷⁵As, so no other isotope can be used. In HCl or HClO₄ the plasma makes ⁴⁰Ar³⁵Cl⁺ (m/z 74.93 against 74.92 for As): a quadrupole cannot resolve them. Use HNO₃ (or collision/reaction-cell…
The full model answer
Worked out step by step, as it should go in the answer booklet. The solution sheet follows it.
- (i) 2 marks: the plasma gas of the torch; its ionisation energy (15.76 eV) exceeds the first IE of nearly all elements (not F, Cl, Ne, He), so analytes give singly charged ions; it is inert, giving no oxide or compound lines; the argon plasma is stable and cheap.
- (ii) 3 marks: arsenic is monoisotopic, ⁷⁵As, so no other isotope can be used. In HCl or HClO₄ the plasma makes ⁴⁰Ar³⁵Cl⁺ (m/z 74.93 against 74.92 for As): a quadrupole cannot resolve them. Use HNO₃ (or collision/reaction-cell mode with He or O₂ mass shift to AsO⁺ at 91).

CoversICP-MSArgon in ICP-MSacids for arsenic
All parts of Question 1
- 1(a) Classification of samples by size5 marks
- 1(b) Stationary-phase film thickness in gas chromatography5 marks
- 1(c) Percentage of amine from the absorbance of its picrate5 marks
- 1(d) Principle of vapour-generation AAS5 marks
- 1(e) Purge gases in thermogravimetric analysis5 marks
- 1(f) X-ray fluorescence in X-ray diffraction5 marks
- 1(g) Argon in ICP-MS; acids for arsenic5 marks
- 1(h) MnO₂ in pyrolusite by back titration5 marks