Q1(b). Particle diameter in chromatography

Question

Why is the diameter of stationary-phase particles critical in chromatography?

Answer in brief

Smaller uniform particles improve efficiency, but require higher pressure.

Key points

  • The plate height is represented by H=A+B/u+Cu. Smaller packing particles reduce multiple-path dispersion, the A term, and shorten mass-transfer distances, reducing C.
  • Consequently H falls and the number of plates N=L/H rises for a fixed column length, producing narrower peaks and improved resolution.
  • Smaller particles can permit faster operation without a large efficiency loss. A narrow particle-size distribution and good packing are also important.
  • Pressure drop approximately varies as ΔP∝ Lu/dₚ². Very small particles therefore demand pressure-rated equipment and good filtration; choose size to balance resolution, time and pressure.

The full model answer

As written for the answer booklet: the diagram carries the working, the lines stay short. Read here only; it is not offered as a download.

Model answer to Q1(b), 2020 Mains Paper III: solution pages 1 and 2
Solution pages 1 and 2 of 20

CoversParticle diameter in chromatographyColumn efficiency