(i) What types of analyte can be separated by HPLC but not by GLC? (ii) What is the range of particle size used as stationary phase packing material in most packed GLC columns?
Answer in brief
(i) Non-volatile, thermally unstable, ionic and high-molar-mass analytes. (ii) 60 to 120 mesh = 0.25 to 0.12 mm (250–120 µm); usual grades 60–80 and 80–100 mesh.
Key points
- GLC: the analyte must be volatile and thermally stable at the column temperature (50 to 300 °C); HPLC needs only that it dissolves in the mobile phase.
- HPLC alone separates non-volatile, thermally labile, ionic and highly polar species and macromolecules (proteins, peptides, nucleic acids, synthetic polymers).
- Packing: a solid support of diatomaceous earth (Chromosorb) coated with 5 to 10 % of the liquid stationary phase.
- Size: 60–80 mesh (0.25–0.18 mm) or 80–100 mesh (0.18–0.15 mm); at the finest 100–120 mesh (0.15–0.12 mm).
- Small, uniform particles give more plates; still smaller particles make the pressure drop across the column too high.
The full model answer
As written for the answer booklet: the diagram carries the working, the lines stay short. Read here only; it is not offered as a download.

CoversHPLCgas-liquid chromatographypacked columnmesh sizeChromosorbnon-volatile analytes
All parts of Question 1
- 1(a) Relative standard deviation of replicate results5 marks
- 1(b) HPLC against GLC; particle size of packed GLC columns5 marks
- 1(c) Sources of radiation for infrared spectrometry5 marks
- 1(d) Graphite-furnace AAS compared with flame AAS5 marks
- 1(e) Calibration of the temperature axis in thermogravimetry5 marks
- 1(f) K and L lines in XRF; wavelength-dispersive spectrum of Zn5 marks
- 1(g) Interference of calcium in the ICP–MS determination of iron5 marks
- 1(h) Type metal: constituents and non-destructive analysis5 marks