Q1(e). Calibration of the temperature axis in thermogravimetry
Question
How is temperature axis in thermogravimetry method calibrated?
Answer in brief
By the Curie-point method: ferromagnetic standards heated in the pan with a small magnet outside the furnace lose their magnetism at their Curie temperatures and give sharp apparent mass steps at known temperatures.
Key points
- Problem: the thermocouple sits near the sample, not in it, so the recorded temperature lags or leads the true sample temperature.
- Curie-point method: the magnetic pull adds an apparent mass; at the Curie temperature the metal turns paramagnetic and a sharp apparent mass step is recorded.
- Several standards cover the range: alumel (about 163 °C), nickel (354 °C), perkalloy (596 °C), iron (780 °C), Hisat-50 (about 1000 °C).
- Fusible-link method: a weight hung on a pure-metal wire drops when the wire melts (In, Sn, Pb, Zn, Al, Ag, Au).
- Calibration is run at the same heating rate, atmosphere and crucible as the analysis; observed against certified values gives the correction curve.
The full model answer
As written for the answer booklet: the diagram carries the working, the lines stay short. Read here only; it is not offered as a download.

Coversthermogravimetrytemperature calibrationCurie pointfusible linkTGA
All parts of Question 1
- 1(a) Relative standard deviation of replicate results5 marks
- 1(b) HPLC against GLC; particle size of packed GLC columns5 marks
- 1(c) Sources of radiation for infrared spectrometry5 marks
- 1(d) Graphite-furnace AAS compared with flame AAS5 marks
- 1(e) Calibration of the temperature axis in thermogravimetry5 marks
- 1(f) K and L lines in XRF; wavelength-dispersive spectrum of Zn5 marks
- 1(g) Interference of calcium in the ICP–MS determination of iron5 marks
- 1(h) Type metal: constituents and non-destructive analysis5 marks