Q1(d). Graphite-furnace AAS compared with flame AAS

Question

What is the advantage and disadvantage of GF-AAS as compared to Flame-AAS?

Answer in brief

Advantage: far higher sensitivity (absolute detection limits 100 to 1000 times lower, picograms) from only 10–50 µL of sample. Disadvantage: poorer precision, slow analysis, stronger background and matrix interference.

Key points

  • GF-AAS: 10–50 µL of sample is dried, ashed and atomised inside an electrically heated graphite tube in argon.
  • Atomisation is nearly 100 % efficient (a flame atomises as little as 0.1 %); absolute detection limits are 100 to 1000 times lower: picograms.
  • Very small sample (microlitres); solids, slurries and viscous samples are taken directly.
  • Poorer precision: about 5–10 % against about 1 % for the flame; slow, one element at a time.
  • Stronger background and matrix effects: background correction (D₂ lamp or Zeeman) and matrix modifiers are needed.